Results

Results Icpms.emf

 

REAGENTS UTILIZED

All chemicals were of analytical grade unless noted otherwise and were used as received. Concentrated nitric acid (HNO3, 70%) and HCl, both trace metal-grade, were used for digestion of aminoboration samples. A 2% HNOsolution prepared using nitric acid certified for trace element analysis and ultra-high purity water (18MΩxcm) were used to prepare all ICP-MS samples.

The two standard stock solutions used are shown in Table 1. Stock A was used in preparing the calibration curve while stock B was used in making the quality control. Single-element standards were used in preparing the internal standards, composed of Sc at 2000 ppb and Ge, Rh, Bi, and In at 400 pb, respectively. Images and Table 1 below show the information for each standard solution.

Table 1: Concentrations in ppm for each component in Stock A and Stock B for calibration curves and quality controls, respectively.

Used Chemicals

 

 

METHOD SEQUENCE

Find below in Table 2 the method sequence used during the ICP-MS operation in sample analysis. Subsequent to running the initial precipitate sample (026SMPL.D#) the concentration was determined to be >200 ppb. Therefore, the following precipitate samples (028-034SMPL.D#) were removed, and diluted 10-fold for analysis at the end of the sequence and found to contain appropriately dilute ppb content. Note also that QC1 and QC2 labels were mistakenly swapped when inputting the sequence, therefore, QC1 corresponds to that of Fe(lower conc) and QC2 corresponds to Fe & Pd.

Table 2: Method sequence for analysis on ICP-MS.

Analytical Sequence

CALIBRATION STANDARDS

The analytes of interest were quantified using a calibration curve generated from multi-elemental standards and as shown in Figure 2, with plots displaying linearity with a high coefficient of determination.

Calibration Curves

 

Figure 2: Calibration curves for measured analytes in varying modes. The fit parameters are summarized in Table 3.

 

 

Table 3: Determination of degree of fit for element calibration curves. The fit equation is given as Y = ax +where Y represents the intensity (cps) and x is the concentration of the analyte (ppb). DL means Detection limit and BEC is the background equivalent concentration.

Mass Element r a b(blank) c DL BEC Unit
56 Fe 0.9990 3.34E-02 5.63E-03 0 8.33E-02 1.68E-01 ppb
56 Fe 0.9990 7.85E-03 1.68E-03 0 4.63E-02 2.14E-01 ppb
60 Ni 0.9991 1.52E-03 1.66E-04 0 6.12E-02 1.09E-01 ppb
105 Pd 0.9990 1.72E-03 1.31E-05 0 6.78E-03 7.60E-03 ppb
107 Ag 0.9996 2.51E-03 1.42E-05 0 6.16E-03 5.67E-03 ppb
109 Ag 0.9981 2.30E-03 1.33E-05 0 8.55E-03 5.77E-03 ppb
206 Pb 0.9991 2.13E-03 1.09E-05 0 5.05E-03 5.14E-03 ppb
207 Pb 0.9990 1.83E-03 7.44E-06 0 4.43E-03 4.07E-03 ppb
208 Pb 0.9991 8.38E-03 3.88E-05 0 5.84E-03 4.63E-03 ppb

 

Analytical figures of Merit

An internal standard is useful in monitoring analyte recoveries in situations of instrumental drift. The choice of internal standards is based on similarities between the mass and ionization energy of elements.  In this study, the internal standard that is used for 56Fe, 60Ni, and 208Pb was 45Sc, 72Ge, and 209Bi respectively while 103Rh monitored 109Ag, and  105Pd.   The recoveries of the internal standards were remarkable as shown in Table 4. Table 5  shows the range of calibration standards that encompassed each analyte, with the concentration rows corresponding to blank, then standards in increasing concentration moving down the column. While the calibration curve accounted for the 56Fe, 60Ni, 109Ag, and 208Pb, most analyte concentrations were exceedingly low and were within the range of the calibration blank and lowest concentration standard. Ideally, calibration curves for these analytes would be expanded slightly at the lower end of the concentration range to capture more of the sample concentrations, excluding the possibility of less dilute samples. Note that Pd content was not analyzed for this series of samples, and therefore the calibration ranges are not bracketed for this analyte. Additionally, the isotope system for Pb is known to be unstable, and therefore, 208Pb is reported as total lead content. 

 

Table 4: Recoveries for internal standards used. Note that [#1] represents Helium mode, [#2]-Hydrogen mode, and [#3]-No gas mode

Internal Standard

 

Table 5: Concentration bracketing for KOH electrolyte samples

Conc Bracketing Koh

 

As shown by the bracketing in Table 6, the 105Pd concentrations of analyzed samples were adequately accounted for by the calibration standards. Furthermore, the dark green background cells indicate that an appropriate range of standards accounted for the median 105Pd content. Therefore, this calibration curve was successful in aiding quantification for these samples. Note that bracketing for other analytes correspond to that of electrolyte samples, and are not representative of the aminoboration samples. 

Table 6: Concentration bracketing for aminoboration samples.

Aminoboration Bracketing

Data Reduction

Note that initial blank runs were excluded from the analysis for LOD. All subsequent blanks were used for this purpose, excluding that of the calibration standard blank.

Electrolytes

Table 6: Electrolyte adjusted analyte concentrations.

Folder: Misc Info: Dil factor ALS Vial: Fe /  56 [#2] Ni /  60 [#3] Pd /  105 [#3] Ag /  107 [#3] Ag /  109 [#3] Pb /  208 [#3]
042SMPL.D# Fe_Pure_dil 0.62569444 3111 355957.839 -19890.924 9457.79979 84397.6037 91160.8026 1279.31247
043SMPL.D# Unpure_dil 0.62847222 3112 355772.487 -22082.798 8083.1509 39099.3963 42408.0804 901.883254
044SMPL.D# 50 CV_dil_1 0.63194444 3201 354654.22 -20974.251 7394.54049 26091.911 29482.4053 1062.54404
045SMPL.D# 100 CV_dil_1 0.63541667 3202 356016.042 -17587.192 4966.42378 21193.635 20296.4745 281.751095
046SMPL.D# 250 CV_dil_1 0.63888889 3203 356828.276 -15589.827 8306.96227 17944.894 17662.9997 743.986956
047SMPL.D# 500 CV_dil_1 0.64236111 3204 356314.23 -24136.726 4817.36838 17192.1616 16222.9869 319.186287
048SMPL.D# 50 CV_dil_2 0.64583333 3205 354643.738 -24537.8 24342.7462 15111.3697 14767.3652 112.13835
049SMPL.D# 100 CV_dil_2 0.64930556 3206 354676.621 -23433.484 23628.5565 18102.6947 17467.8236 -518.8919
050SMPL.D# 250 CV_dil_2 0.65277778 3207 351898.887 -278.7743 22268.1616 21398.9713 21567.8828 -585.55975
051SMPL.D# 250 CV_dil_3 0.65625 3208 356047.007 -23815.984 24688.2994 12707.3177 14138.6266 932.131064
052SMPL.D# 500 CV_dil_2 0.65972222 3209 332098.66 -25731.004 21875.3387 12838.9342 12161.4529 -232.76795
054SMPL.D# P_50 CV_dil_1 0.66597222 3210 352489.934 -16052.392 24628.4717 14113.697 14152.4709 600.290358
055SMPL.D# P_100 CV_dil_1 0.66944444 3211 354817.77 -24510.812 20912.9593 18301.5006 20472.9853 1120.1597
056SMPL.D# P_500 CV_dil_1 0.67291667 3212 355014.542 -24556.356 19834.6625 12617.2168 11225.5598 -48.459485
057SMPL.D# YJS Purified dil 0.67638889 3301 354405.267 -24865.074 22022.7433 15044.5036 12857.8231 -51.530526

Aminoboration

Table 7 shows the adjusted concentrations with the utilized dilution factors for aminoboration samples. It is clearly evident that the majority of Pd content was present in the precipitate samples, making the higher dilution factor necessary for their analysis. 

Table 7: Adjusted concentrations based on dilution factors for aminboration samples.

Folder: Misc Info: Date: ALS Vial: Dil factor Pd /  105 [#3] ppb
021SMPL.D# EGM_1_SUP 12/1/23 3101 3500.309 2846.451
022SMPL.D# EGM_2_SUP 12/1/23 3102 3522.378 27671.8
023SMPL.D# EGM_3_SUP 12/1/23 3103 3619.24 46724.38
024SMPL.D# EGM_4_SUP 12/1/23 3104 3571.783 22895.13
025SMPL.D# EGM_5_SUP 12/1/23 3105 3573.289 10755.6
064SMPL.D# EGM_1_PPT (10x more) 12/1/23 3302 35008.72 822354.8
066SMPL.D# EGM_2_PPT (10x more) 12/1/23 3303 35330.96 1091373
068SMPL.D# EGM_3_PPT (10x more) 12/1/23 3304 35537.58 999672.2
070SMPL.D# EGM_4_PPT (10x more) 12/1/23 3305 35566.6 1069488
072SMPL.D# EGM_5_PPT (10x more) 12/1/23 3306 35666.65 1026486