Methodology

 

Sample Selection

A dataset of ten unionid bivalve shells collected across the Wichita River Watershed was reviewed based on completeness, and the TDS of the water at their sample location. To show the eastward salinity gradient of the watershed contributed by Permian evaporites west of the sampling area, three shells were selected from Lake Kemp, Holiday Creek at Wichita Falls, and and eastern portion of the Wichita River river reaches with high, intermediate, and low TDS in water at the collection locations. A fourth specimen was selected from the Little Wichita River, which gives insight into a nearby environment with different water sources. This set of of specimens was useful for comparison of water composition reflected in carbonate composition across the WRW system.

Wrw Water Tds

Figure 1. Wichita River Watershed TDS of Water at Bivalve Collection Localities

Wrw Shell Sr

Figure 2. Wichita River Watershed Bivalve 87Sr/86Sr

Sample Preparation

Following selection, shells were cut longitudinally from umbo to ventral margin using an Isomet saw with water cooling to minimize damage to sample and equipment. The sectioned halves were mounted in epoxy resin containing an indium tracer and left to cure for ~3 days. Embedded specimens were trimmed to form a billet of ~2 mm thickness and polished on a rotating wheel using a progressive grit sequence from 20 μm to 0.5 μm. Polishing helped to produce a smooth surface suitable for laser ablation.

Lwr02.1 012        Img 3861      Img 3867      Img 4475

Figure 2. From Left to Right: Specimen LWR021-012 (Uncut), Specimen LWR021-012 (Cut transect), Specimens Curing in Epoxy Resin Mount, Final Cut and Polished Billets for Analysis

 

Laser Ablation Setup

Elemental data of shell transects was collected using an NWR193 excimer laser ablation system. Ablation conditions were selected using Agilent MassHunter suited to to carbonate material and to allow for high resolution data from growth layers of shells.

Spot geometry 10×75 μm rectangular aperture
Scan speed 5 μm/s
Fluence 3.8 J/cm²
Repetition rate 10 Hz
Carrier gas flows He = 0.80 L/min; Ar = 0.80 L/min
Sampling depth 7 mm
Sampling period 0.347 s per cycle

Before data on shell transects were collected, pre ablation at a reduced energy was used to remove surface contamination. Three consecutive analytical sessions were performed. The first was terminated early for NIST610 upon noticing the equipment was focused on the bottom of the glass slide. The final session was used to collect reading for final standard transect of NIST610, which timed out during the second session

 

ICP-MS Instrumentation and Selected Analytes

Elemental measurements were collected using an Agilent 8900 ICP-MS/MS run entirely in no gas mode, with all analytes measured at a 20 ms dwell time. The analytes were selected to capture variability relating to reconstructing seasonal shifts in water chemistry, TDS, redox effects, and human pollution in the Wichita River Watershed. Calcium 43Ca was used as the internal standard for concentration normalization with an assumed 40.0432 wt%. The table below summarizes possible uses of data collected on each analyte.

Salinity and Water Chemistry (TDS) Na23, Sr88, Ba137/Ba138
Temperature/Seasonality Mg24/Mg25 (Uses Mg/Ca)
Redox/Runoff Mn55, Fe57
Internal Standard/Proxy Ratios Ca43, Ca44
Anthropogenic Pollution Pb208
Other Zn66/Zn68, Se77/Se82

 

Analytical Sequence

In transects of bivalve shells as unknown, ordering from 1-4 is determined by transect proximity to umbo of shell, with 1 being most proximal to umbo and 4 being most distal to umbo. Transects were collected from the periostracum of the shell’s outer surface, to the inner surface of the shell.

  1. MACS-3: 1-3

  2. NIST 612: 1-3

  3. NIST 610: 1-3

  4. Unknowns:
    WR01-006-04
    WR01-006-01
    WR01-006-02
    WR01-006-03
    HC02.1-009-01
    HC02.1-009-02
    HC02.1-009-03
    HC02.1-009-04
    LK01-22-04
    LK01-22-03
    LK01-22-02
    LK01-22-01
    LK021-012-01
    LK021-012-02
    LK021-012-03
    LK021-012-04

  5. MACS-3: 4-6

  6. NIST 612: 4-6

  7. NIST 610: 4-6

Screenshot (23)

Figure 3. Ablation Transect of Shell Perpendicular to Growth Layers

 

Iolite Reduction

Laser ablation data was processed using Iolite 4 set to “Trace Elements” for the data reduction scheme. Laser log files assigned time intervals to the 3 ablation datasets in this timeseries. The first was terminated early for NIST610 upon noticing the equipment was focused on the bottom of the glass slide, while the second timed out while recording final NIST610 standard and had to be repeated. Baselines were automatically selected by iolite and manually verified. Spline baselines were assigned to the unknowns, linear baselines were assigned to the standards.

All concentrations were internally standardized to 43Ca, assuming 40.0432 wt% Ca. MACS-3 served as the primary calibration standard, with NIST 610 and NIST 612 used as secondary quality control materials. NIST610-1 was excluded due to low intensities across all analytes, and NIST612 yielded unreliable readings due to focus on bottom of slide during data collection. Sensitivity drift was corrected by linear interpolation by bracketing standard blocks. After this, the concentrations were exported as a timeseries for each shell transect.

Untitled 2 (2)

Figure 4. Timeseries as displayed in Iolite 4. Gaps visible between 3 stitched datasets in timeseries. Omitted NIST610 bracket notably falls below line outlined by other brackets.

 

Quality Control and Measurement Quality

Performance of ablation was assessed using measurements of NIST 610, NIST 612, and MACS-3.

1. Reduced through the MACS-3 calibration curve.

  • NIST610 – Recoveries had an average of 88% for major elements, excluding considerable bias in Fe (40Ca17O and 40Ar17O interference) and Pb.
  • NIST612 – Recoveries were low across set of analytes, typically falling below 40%. Likely, this is the result of misfocusing during data acquisition.

QC1: MACS-3

 

Analyte RefValue MeanMeasured Recovery 2SE LOD n
Li7 62.9 62.9 1 3.07 0.08 6
Na23 5850 5849.99 1 620.44 0.17 6
Mg24 1720 1720 1 64.73 0.04 6
Mg25 1720 1720 1 67.42 0.16 6
Ca44 376900 376900.06 1 11675.32 15.75 6
Mn55 512 512 1 16.44 0.11 6
Fe57 10500 10500 1 553.37 1.34 6
Zn66 124 124 1 6.43 0.09 6
Zn68 124 124 1 6.71 0.6 6
Sr88 6640 6639.96 0.99999 303.05 0.01 6
Ba137 59.6 59.6 1 3.1 0.01 6
Ba138 59.6 59.6 1 3.15 0 6
Pb208 59.7 59.7 1 5.59 0.02 6

QC2: NIST610

Analyte RefValue MeanMeasured (ppm) Recovery 2SE LOD n
Li7 468 422.51 0.9 9.93 0.15 5
Na23 99415 82390.36 0.83 2023.98 0.63 5
Mg24 432 439.02 1.02 8.93 0.07 5
Mg25 432 449.28 1.04 13.29 0.45 5
Ca44 82144 82411.54 1 2212.5 30.4 5
Mn55 444 395.04 0.89 7.89 0.2 5
Fe57 458 294.45 0.64 8.08 2.35 5
Zn66 460 372.52 0.81 9.3 0.13 5
Zn68 460 388.59 0.84 8.08 1.23 5
Sr88 515.5 435.87 0.85 11.99 0.04 5
Ba137 452 344.25 0.76 7.89 0.02 5
Ba138 452 344.12 0.76 7.98 0.01 5
Pb208 426 254.74 0.6 6.33 0.02 5

QC3: NIST612 (Misfocused Readings)

Analyte RefValue MeanMeasured Recovery 2SE LOD n
Li7 40.2 1.52 0.04 0.06 0.11 5
Na23 103858 81845.81 0.79 883.25 0.17 5
Mg24 68 27.31 0.4 0.36 0.17 5
Mg25 68 27.02 0.4 0.89 0.54 5
Ca44 85002 85390.2 1 982.65 1330.78 5
Mn55 38.7 1.14 0.03 0.07 0.16 5
Fe57 51 12.75 0.25 1.03 2.03 5
Zn66 39.1 2.03 0.05 0.16 0.15 5
Zn68 39.1 6.78 0.17 0.48 1.09 5
Sr88 78.4 39.52 0.5 0.5 0.84 5
Ba137 39.3 2.45 0.06 0.1 0.01 5
Ba138 39.3 2.47 0.06 0.05 0.06 5
Pb208 38.57 1.4 0.04 0.04 0.05 5

 

2. Reduced through the NIST610 calibration curve.

  •  MACS-3 – Recoveries had an average of 117% for major elements, excluding considerable positive bias in Fe (40Ca17O and 40Ar17O interference) and Pb.
  • NIST612   Recoveries were low across set of analytes, typically falling below 40%. Likely, this is the result of misfocusing during data acquisition.

QC1: NIST610

Analyte RefValue MeanMeasured Recovery 2SE LOD n
Li7 468 468 1 10.99 0.18 5
Na23 99415 99415.05 1 2423.02 0.84 5
Mg24 432 432 1 8.73 0.07 5
Mg25 432 432 1 12.74 0.45 5
Ca44 82144 82144 1 2202.44 31.55 5
Mn55 444 444 1 8.87 0.23 5
Fe57 458 458 1 12.52 3.92 5
Zn66 460 460 1 11.47 0.17 5
Zn68 460 460 1 9.56 1.51 5
Se77 138 138 1 5.12 2.04 5
Se82 138 138 1 4.53 6 5
Sr88 515.5 515.5 1 13 0.05 5
Ba137 452 452 1 10.34 0.03 5
Ba138 452 452 1 10.45 0.01 5
Pb208 426 426 1 10.53 0.04 5

QC2: MACS-3

Analyte RefValue MeanMeasured Recovery 2SE LOD n
Li7 62.9 69.26 1.101113 3.38 0.1 6
Na23 5850 6804.17 1.163106 757.25 0.22 6
Mg24 1720 1661.2 0.965814 62.91 0.04 6
Mg25 1720 1629.56 0.947419 64.13 0.16 6
Ca44 376900 373025.3 0.98972 11568.39 16.3 6
Mn55 512 574.22 1.121523 18.44 0.13 6
Fe57 10500 15997.22 1.523545 843.44 2.24 6
Zn66 124 151.83 1.224435 7.88 0.12 6
Zn68 124 146.73 1.183306 7.94 0.75 6
Se77 59.6 5.86 0.098322 1.06 0.97 6
Se82 59.6 5.68 0.095302 1.68 3.36 6
Sr88 6640 7840.65 1.180821 357.91 0.01 6
Ba137 59.6 77.55 1.301174 4.04 0.01 6
Ba138 59.6 77.12 1.29396 4.1 0.01 6
Pb208 59.7 98.13 1.643719 9.25 0.03 6

QC3: NIST612 (Misfocused Readings)

Analyte RefValue MeanMeasured Recovery 2SE LOD n
Li7 40.2 1.67 0.0416 0.06 0.11 6
Na23 103858 96835.66 0.9324 1121.91 0.17 6
Mg24 68 26.58 0.3909 0.44 0.17 6
Mg25 68 25.69 0.3778 0.93 0.54 6
Ca44 85002 84660.78 0.996 910.67 1330.78 6
Mn55 38.7 1.29 0.0334 0.08 0.16 6
Fe57 51 1156.8 22.683 2778.27 2.03 6
Zn66 39.1 19.71 0.5042 0.96 0.15 6
Zn68 39.1 3.34 0.0854 0.87 1.09 6
Sr88 78.4 2.48 0.0316 0.14 0.84 6
Ba137 39.3 7.97 0.2028 0.26 0.01 6
Ba138 39.3 1.29 0.0328 0.57 0.06 6
Pb208 38.57 5.57 0.1445 1.52 0.05 6